Home / Services / Maintenance and calibration
Valve test bench maintenance and calibration: what drifts, what a visit covers, how it is signed off
A valve test bench drifts in three places only: the sealing and clamping parts that meet the valve, the pressure indication across the two gauge ranges, and the hydraulic and water circuits that feed them. This service holds all three inside the record your customers audit — on the interval your own quality procedure sets, with as-found and as-left readings for every instrument touched.
Send the model and last calibration date Email the bench details
Quick answer
| Who it is for | Plants running a bench — ours or another make — whose per-valve records have to survive a customer or ISO audit |
|---|---|
| Covered | Seal discs, jaw pads, blind flange faces; both gauge ranges; transmitter; oil, filter and pump; water and air circuit; interlock function test |
| Interval | Set by your quality procedure. The visit content on this page is what fills that interval |
| Left behind | Date, reference instrument, as-found and as-left per channel, pass or fail per safety device |
| Cannot do | Judge a valve, move the pressure envelope, stand in for an accredited laboratory, rebuild a worn machine |
| Quoted | After the model, the DN range and the date of the last calibration are known |
Three things drift, and only three
A bench is a frame, a clamp, a pump and a set of instruments, and each of the four fails in its own way. Nothing on the list is exotic: a sealing disc flattens, a jaw pad wears a groove, a gauge loses its zero, oil takes on water. What makes them worth a scheduled visit rather than a call when the machine stops is that all four fail quietly — the bench keeps producing a number, and the number is wrong before anyone notices.
| What drifts | Where it shows first | What the visit has to leave behind |
|---|---|---|
| Sealing and clamping parts | The same valve passes in the morning and fails after lunch; the hold falls away faster between runs; a seal face carries a flat or a rolled edge | New discs and pads where the wear list says so, plus a hold record at 1.5× for shell and 1.1× for seat on a class you nominate |
| Pressure indication | The two ranges disagree at a common point; a low-pressure test reads in the bottom sliver of the high-range dial | As-found and as-left per range, with the reference instrument named on the record |
| Hydraulic and water circuit | Build-up gets slower over months; oil turns dark or milky; the filter shows a pressure drop; the filling pump sounds rougher than it did | Oil and filter record, water circuit flush, pump and relief observation written down rather than remembered |
| Safety devices | Not usually visible at all: a guard switch that no longer breaks the clamp circuit, an E-stop wired to stop the pump but not the clamp | Pass or fail per device, tested one at a time, with settings recorded as found |
Why two gauges mean two calibrations
The two-gauge layout exists because one dial cannot serve both ends of the range. A bench built for DN15 to DN600 with water up to 40 MPa carries a low range — 0–4 MPa — and a high range, 0–40 MPa. Small valves are judged on the low range: a DN50 PN16 seat test at 1.1× sits at about 1.76 MPa, and on a 0–40 MPa dial that reading lives in the bottom five per cent of the scale where the pointer has the least room to be read.
The arithmetic is worth doing once. An instrument of 1.6% full-scale accuracy on a 0–40 MPa dial is allowed about ±0.64 MPa anywhere on that dial. At a test pressure of 1.76 MPa that tolerance is roughly a third of the reading — which means the same valve can be called sound or leaking by the instrument alone. On a 0–4 MPa dial the same accuracy class is allowed about ±0.064 MPa, and the question becomes answerable. This is arithmetic, not a measurement claim: it is the reason the standard a valve is sold to decides which range the reading must come from, and it is the reason both ranges are calibrated at every visit rather than the high one alone.
Where a transmitter feeds the HMI or the data export module, it is a third channel and it is checked against the gauge at common points, not against itself. A screen that agrees with a drifting gauge is two wrong numbers agreeing.
What a visit covers
Six blocks, in the order the work is done. Every block ends in a line on the record, and a block that cannot be completed is written down as not completed rather than left blank — that is what makes the record usable in an audit.
A — Sealing and clamping
Discs and pads lifted out and inspected for flat, rolled edges, grooves and hardness; blind flange faces checked for scoring and for a matching seat angle; clamp stroke and jaw alignment checked across the DN range in use, not at one size; clamp cylinder watched for creep under load. Anything on the wear list that is due and not in your stores goes onto the next-visit list with the DN segment it belongs to, so the part can be ordered before it stops the bench.
B — Pressure indication
Both ranges checked at agreed points against a reference instrument, as found and as left; the transmitter compared to the gauge it is supposed to mirror; relief valve observed and its setting recorded as found. Adjusting a setting is a separate written instruction from the plant — the visit does not quietly re-set a safety device to make a number look better.
C — Hydraulic circuit
Oil level and condition, filter element, pump behaviour at build-up, hose and coupling condition, quick couplings for weeping. Oil that has gone milky is water ingress and is a finding in its own right, not a topping-up job: the water came from somewhere, usually the test circuit.
D — Water and air circuit
Tank, level control and drain checked; strainer cleaned; on water-and-air machines such as the SYJ-600SQ, the air side gets its own line: filter-regulator-lubricator, air hose and air-side seals, which dry out faster than the water side because they meet air rather than water.
E — Safety devices
Guard and light curtain, E-stop, clamp interlock and the bench relief are tested one at a time and recorded pass or fail. The test is functional: does the clamp circuit actually break when the guard opens, does the pump actually stop on the E-stop. A device that is present but not wired into the circuit fails the same test as one that is missing.
F — Records
The visit report, the as-found and as-left table, the wear list for the next interval, the oil and filter record and the safety-device table. Where an export module is fitted the same content is left as CSV as well as paper, so the plant can file it in whatever system it already audits against.
The interval belongs to your procedure, not to this page
A supplier who prints an interval in a web page is answering a question that belongs to the customer's quality system: what the plant has to prove, to whom, and how often. What we can do is fill whatever interval is set, with content that covers the four failure modes above. The three tiers below are the template agreed at the first visit — the middle tier is the one whose interval you write.
| Tier | Who does it | Content |
|---|---|---|
| Before each shift | Operator | Gauges at zero with the bench unloaded; clamp travel free; water level; no weeping at couplings; guard in place. Written on a shift sheet, not on memory |
| Periodic — interval set by your quality procedure | Plant maintenance, with our checklist | Discs and pads inspected against the wear symptoms; filter checked; oil level and colour; both gauge ranges compared at one common point; interlock spot check |
| Scheduled service visit | Our technician | Blocks A to F above, with as-found and as-left records and a signed report |
The shift and periodic tiers exist so that the scheduled visit is not the first time anyone has looked. A bench that arrives at a service visit with three months of unreported weeping is a repair job, not a calibration job — and it is quoted as such.
Nine models, and what each wears first
The visit is the same six blocks on every machine; the machine decides where the hours go. The envelopes below are the published ones — the bench overview carries the full specification for each model.
| Model | Layout and envelope | What a visit touches first |
|---|---|---|
| SYJ-150 | Vertical, DN15–DN150, 0–40 MPa, 30 t | Seal discs in the small DN segments, which is where the whole workload sits; the low range 0–4 MPa carries most readings, so it is the range that matters |
| ZSYJ-300A | Vertical, DN50–DN300, 1.6–40 MPa, 80 t | Discs in the DN50–150 band wear fastest; both ranges in daily use, so both are calibrated at every visit |
| SYJ-600SQ | Vertical, DN50–DN600, water 40 MPa / air 4 MPa, 90 t | Two media, two sets of seals: air-side seals dry and harden, water side scales; air FRL and water strainer both on the list |
| YSYJ-300 | Vertical, DN50–DN400, water 40 MPa / air 0.8 MPa, 80 t | Air circuit at 0.8 MPa is a low-pressure check: the low range gauge and the air FRL decide whether the result means anything |
| SYJ2-300-600 | Two stations, DN50–300 and DN350–600, PN6–PN100 | Two clamps, two plate sets: each station holds on its own, so each station is proved on its own record, not one for the pair |
| WBD-600 | Horizontal, ram and three jaws, DN50–DN600, 0–16 MPa, 150 t | Jaw pads and jaw faces, because a horizontal table takes the valve's weight on the jaws; water pooling on the table after each shift |
| SYJ-600M | Vertical clamp, DN350–DN600, 0–10 MPa, 126 t | Large plate flatness and lifting gear; a plate that is no longer flat leaks at the rim and looks like a seat failure |
| DFS2-600 | Two stations, DN50–300 (grip 80 t) and DN350–600 (grip 126 t), 0–7 MPa, USB export | Both stations' clamps plus the export module: a record that stops exporting is a failed audit before it is a machine fault |
| ZZC2-200 | Automatic gate valve line, two stations testing together, DN50–200, PN5 MPa, printed record | Cycle count is the wear driver: pads and discs go first, then the printer and the record path; a line that runs all shift has no slack for a soft seal |
A scheduled visit against waiting for the breakdown
The comparison is not about cost per hour, because nobody can price a stopped line honestly. It is about which of the two leaves evidence.
| Call when it stops | Scheduled visit | |
|---|---|---|
| When the fault is found | After a valve has been passed or failed on a wrong reading | When the instrument is out of tolerance, not when the customer complains |
| Parts | Ordered after the machine is already down, in whatever DN segment turns out to be needed | Wear list from the previous visit, so the part is in stores before it is due |
| Records for an audit | Assembled afterwards from memory and photographs | As-found and as-left, dated, on file |
| Scope creep | A seal change becomes a pump, a cylinder and a plate set because everything is found at once | One block at a time, so wear is visible as a trend rather than a shock |
| Who decides | The breakdown decides | The plant decides, inside the window it chose |
Window, downtime and parts
Two stages, and the first one is why the second one does not overrun. First a survey: bench model, DN range, media, control system, the last calibration date, and photographs of the seal faces and the cabinet. From that, the workshop returns the visit plan — which blocks apply, which parts are likely due, and which of those are on the shelf against made to order. Then the visit itself is planned into a window the plant picks, and the plant knows before the day which checks need the bench pressurised and which can be done around production.
Checks that need pressure — the hold record, the gauge comparison at a working pressure, the interlock test under load — are grouped so the bench is out of service once rather than three times. Nothing is pressurised while a person is at the clamp; the clamp is proved empty and the guard closed before any block that involves pressure begins. If a part turns out to be made to order rather than on the shelf, the block it belongs to moves to the next window instead of holding the whole visit open — that choice is written into the report, not left to be discovered later.
How a visit is accepted
- Both ranges recorded. As-found and as-left for 0–4 MPa and 0–40 MPa, at the agreed points, with the reference instrument named on the record.
- A hold the plant nominated. Shell at 1.5× and seat at 1.1× on a DN and class the plant chooses, with the pressure at start and end of the hold written down.
- Repeatability stated, not implied. The same valve run three times, the three results printed side by side, and the spread named rather than described as good.
- Safety devices pass or fail, one by one. Guard, E-stop, clamp interlock and relief each tested and recorded; a device that cannot be tested is recorded as not tested, with the reason.
- Wear list for the next interval. Parts that are due, the DN segment each belongs to, and whether each is stock or made to order.
- Records handed over, not kept. Signed report plus CSV where an export module is fitted, in the plant's own filing structure.
What this service is not.
— It does not judge a valve or issue a pass certificate. The record shows what was measured; the standard and the operator still decide, exactly as on the test procedure.
— It does not move the envelope. Class and bore are set by pump, seals and frame; neither calibration nor a new seal disc changes them, which is the same rule that applies to a retrofit.
— It does not replace an accredited laboratory certificate. Where the plant needs accreditation, that certificate comes from the plant's own external laboratory; the visit record supports it and does not imitate it.
— It does not rebuild a worn machine. If the frame, the cylinder or the pump needs replacing, that is the rebuild scope, quoted as a rebuild.
— It does not re-set a safety device without a written instruction, and it does not leave a setting changed without saying so.
Questions that come with the first enquiry
Do you service benches from other manufacturers?
Yes for the measurement and wear side: gauges and transmitters, seal discs, jaw pads, blind flange faces, hydraulic and water circuits — as long as the parts can be matched by nameplate or by measuring the old part. Control work on an unknown system is surveyed before it is promised: if the programme cannot be read and backed up, that is said before the visit rather than discovered during it. The same rule applies to the retrofit, which is why that page lists five feasibility inputs.
How often should a valve test bench be calibrated?
The interval belongs to your quality procedure — what you have to prove, to whom, and how often. What this service does is fill whatever interval you set and leave a record showing the date, the reference, the as-found value and the as-left value. If the procedure has no interval yet, the three-tier template above is agreed at the first visit and written into the report so the next date is not a guess.
Can the bench keep running during the visit?
Partly. Functional checks, seal inspection, oil and filter, water circuit and paperwork can be done around shifts. The blocks that need pressure — the hold record, the gauge comparison at a working pressure, the interlock test under load — need the bench out of service for the agreed window. Those blocks are grouped on purpose, and the window is set by the plant before the day.
Is your calibration record the same as an accredited certificate?
No, and it does not pretend to be. The visit record names the reference instrument used, shows as-found and as-left, and is signed and dated; where the plant's customer requires a certificate from an accredited laboratory, that certificate is issued by the plant's own external laboratory. A page that offers to replace it would be selling something it cannot deliver.
The bench loses pressure between runs — will a visit fix that?
Sometimes, and the record says which. If the loss comes from a flattened disc, a grooved jaw pad or a scored flange face, it is a wear part and the visit fixes it. If it comes from the pump, the relief, a cylinder seal or the pipework, it is a repair and is quoted as one — the visit finds it and writes it down rather than changing the topic. Reading the leak on the wrong range is a third possibility, and that one is a calibration finding.
What should we have ready before the visit?
The machine model and the DN range in use, the date and result of the last calibration, the quality procedure clause that sets the interval, the parts currently in stores, and photographs of the seal faces and the inside of the cabinet. With those, the workshop can say before the day which blocks apply and which parts are likely due — and whether any of them is made to order. Send them through the quote form and the visit plan comes back within 48 hours.
The bench — ours or another make — the DN range it works, the media, and the interval your procedure sets. The workshop comes back within 48 hours with the visit plan, the blocks that apply, and the parts that should already be on your shelf.